51% (89) 1.9k views 9 pages

ABB Sensi+ Manual Laser Alignment Service Guide

AI-enhanced description

This service guide provides a manual laser alignment procedure for ABB Sensi+ instruments when the Autotune feature fails due to poor spectroscopic fit quality, often caused by laser temperature shifts. The procedure involves adjusting the laser TEC temperature setpoint within a specified range while monitoring fit quality and absorption peaks, using a compliant gas blend such as pure CH4.

Uploaded by Valentina42
Download
/ 9